Convert flow-cup efflux time to kinematic viscosity and, when density is provided, dynamic viscosity, using the equations in ISO 2431:2011. Select the flow cup used for your measurement.
Active Conversion Equation — ISO 2431 Cup No. 4
Select ISO 2431 Flow Cup
Kinematic Viscosity
Dynamic Viscosity
Practical Note
Select a flow cup that gives an efflux time between 30 and 100 seconds (ISO 2431 §7.2.1). Condition the sample and cup to 23.0 ± 0.5 °C before testing. This method is only reproducible for Newtonian or near-Newtonian liquids — if your material contains thixotropic or flow-arresting additives, treat the result as a consistency check, not a true viscosity, and verify with a rotational viscometer (ISO 3219).
Conversion equations are taken directly from ISO 2431:2011, Table 1 (ν = A·t − B/t, valid for 30 s ≤ t ≤ 100 s per cup).
Dynamic viscosity (cP) = Kinematic viscosity (cSt) × Density (g/cm³)
Built by PolymerDNA — more formulation references in the Formulae Masterbook.